HEARTSTART FRx 861304 OWNER’S MANUAL
AP
P
E
ND
IC
ES
Patient treatment protocol options
parameter
settings
default
default description
“call EMS” voice
reminder timing
• At power on (when the 
user turns on the FRx)
• At power on and at the 
start of the first patient 
care pause
• At the start of the first 
patient care pause 
• No  reminder
At the start of 
the first 
patient care 
pause.
Provides a voice reminder to 
make sure emergency medical 
services have been called, at 
the start of the first patient 
care pause.
shock series
1, 2, 3, 4
1
The automatic protocol pause 
for CPR is activated each time 
a shock is delivered.*
During the protocol pause, 
the FRx does not perform 
rhythm analysis.
The length of the protocol 
pause after a shock series is 
completed is determined by 
the protocol pause timer 
setting.
shock series interval
(minutes)
1.0, 2.0,
∞ 
(infinity)
1.0 
A delivered shock must occur 
within 1 minute of the 
previous shock to be counted 
as part of the current shock 
series.
NOTE: This parameter is only 
applicable when the shock series 
is not configured to the default 1 
shock.
* A shock series begins when a shock is delivered after the FRx is turned on. A new shock series begins after a protocol 
pause. If shock series is configured for 2 or more, a new shock series also begins if the time since the previous shock 
exceeds the shock series interval setting.