Stratos LV/LV-T Technical Manual  107 
N
OTE
The attending physician must notify the BIOTRONIK Service 
Center about which of these events he/she wishes to be 
informed. 
2.17.4.3 Patient Report 
It is possible to trigger a transmission through magnet 
application over the CRT-P.  The attending physician must 
inform the patient in detail about operating the device and about 
the physical symptoms which would warrant a magnet 
application by the patient. 
C
AUTION
 
Patient-Activated Report - The magnet effect must be 
programmed “synchronous” if the [Patient Report] function is 
activated. 
2.17.5 Description of Transmitted Data 
The following data are transmitted by the Home Monitoring 
system, when activated.  In addition to the medical data, the 
serial number of the implant is also transmitted. 
The Monitoring Interval 
Type of Last Home Monitoring Message and Time of the 
Transmission of the last Home Monitoring Message provide 
information regarding the interval of Home Monitoring data. 
Resynchronization Therapy 
•  % Pacing in the Atrium / 24 hours [%] 
•  Atrial intrinsic rhythm [%] 
•  CRT - %ventricular pacing [%] 
•  Ventricular paces (AV Delay expired) [%] 
•  Ventricular paces (triggered by RV Sense) [%]