SECTION 4—LIFTING THE PATIENT
15
FIGURE 4.2   Attaching a Sling to the Swivel Bar
Attaching a Sling to the Optional Cradle Attachment
 WARNING
The maximum weight capacity of the Jasmine Patient Lift is 500 lbs. DO NOT exceed the 
maximum weight capacity of the patient lift otherwise, injury or damage may occur.
Before using the sling, read and understand all operating instructions provided in the sling 
owner’s manual.
Bleached, torn, cut, frayed or broken slings are unsafe and could result in injury. Discard 
IMMEDIATELY. DO NOT alter slings in any way.
Invacare slings are made specifically for use with Invacare Patient Lifts. For the safety of the 
patient, DO NOT use slings and patient lifts of different manufacturers.
NOTE: For this procedure, refer to FIGURE 4.3 on page 16.
1.
Place the straps of the sling over the pegs on the cradle attachment. 
2.
Match the corresponding colors on each side of the sling for an even lift of the patient.
NOTE: The Invacare Cradle Attachment has four hookups.
 WARNING
The patient lift MUST be moved with the push handles on the mast assembly. DO NOT use the 
cradle attachment handle to move the patient lift - otherwise, injury or damage may occur.
DO NOT move the patient outside the center of gravity achieved after the patient is suspended 
in the sling and the patient lift adjusts.
Sling straps MUST be fully 
engaged on swivel bar hooks
Sling Strap