A&D HV-200KV-WP Manuale Utente

Pagina di 72
HV-WP/HW-WP Series
Page 17
7. Basic Operation
7.2.
7.2.
7.2.
7.2.  Tare (And Net Display)
Tare (And Net Display)
Tare (And Net Display)
Tare (And Net Display)
"Tare" is used to cancel the mass of a container, receptacle, case, bag, etc. which is
put on the weighing pan to contain the item to be weighed.
Caution
Using a tare value reduces the weighing range.
The current tare value is reset by pressing the 
 ZERO 
 switch or turning the
scale off. (Reset value is zero.)
The preset tare value must be within the minimum weighing range for the
HV-WP series.
7.2.1.
7.2.1.
7.2.1.
7.2.1. The Way of Tare Input by Weighing
The Way of Tare Input by Weighing
The Way of Tare Input by Weighing
The Way of Tare Input by Weighing
Step 1 Put the container item on the weighing pan.
Step 2 Press the 
 TARE 
 switch. The display becomes zero and the net mark is displayed.
Step 3 It is now possible to put something into the container and to read its net display.
Step 4 Remove all things on the weighing pan.
7.2.2.
7.2.2.
7.2.2.
7.2.2. The Way of Digital Input (Preset Tare)
The Way of Digital Input (Preset Tare)
The Way of Digital Input (Preset Tare)
The Way of Digital Input (Preset Tare)
Step 1 Press and hold the 
 SET 
 switch and press the 
 TARE 
 switch.
Then the blank or stored tare value is displayed. A blank display means that the
tare value is zero (reset value), and   or 
 blinks.
Step 2 Set the preset tare value by using the following switches.
 ∧ 
 switch
Selecting the number of the figure.
 < 
 switch
Selecting a figure.
 F 
 switch
Selecting the polarity (+,-).
Step 3 Press the 
 ENTER 
 switch to store the new preset tare value.
Then the scale displays a net value with the tare value subtracted from the gross
weighing value.
Step 4 It is then possible to put something into the container and to read its net.
Step 5 Remove all things from the weighing pan.