Health O Meter 498KL Manuale Utente

Pagina di 20
 
 
TROUBLESHOOTING  
 
 
 
 
 
 
 
 
 
498KL
 
 
Refer to the following instructions to check and correct any failure before contacting service personnel. 
 
 
SYMPTOM 
 
 
POSSIBLE CAUSE 
 
 
 
CORRECTIVE ACTION 
Scale does not turn on 
1.  Battery problem 
2.  Bad Power Supply or Not Plugged In 
3.  Electrical Outlet not functioning 
1.  Check / Replace the batteries 
2.  Check / Replace Power Supply Adapter 
3.  Use a different outlet 
 “LobAt” 
shows in display 
The batteries are low 
Replace batteries according to instructions 
Questionable weight or 
the scale does not zero 
1.  External object interfering with the scale 
2.  The display did not show “0.0” before 
weighing 
3.  Scale is not placed on a level surface 
4.  Scale is out of calibration 
1.  Remove interfering object from the scale 
2.  Ask the patient to step off the scale. Zero the 
scale and begin weighing process again. 
3.  Place scale on level surface 
4.  Check weight with know weight value 
 “OVER” 
shows in display 
Overload.  The load on the scale exceeds scale 
capacity (500 lbs / 220 kg) 
Remove weight, turn scale off, turn scale on again 
and try again 
 “UNDER” 
shows in display 
Negative weight, indicating a tare was present. 
Load on the scale is below initial zero point 
Clear scale of any weight and re-zero scale by 
pressing ZERO button 
 
CALIBRATION    
 
 
 
 
 
 
 
 
 
          
 
 
This scale has been factory calibrated and does not require calibration prior to use. 
 
The calibration procedure may be performed using a known weight in either pounds or kilograms. Simply set the scale for the desired 
unit of measure using the lb/kg key and start the calibration procedure. The calibration is performed using lb or kg, according to the unit 
of measure used upon entering into the calibration path, meaning the scale will default to either lb or kg, if that was the setting left on 
the scale for weighing.  Note:  Calibration weight must be no less than 100 lb (40 kg) and no more than 500 lb (200 kg).  
 
 
 
 
 
 
 
ACTION 
DISPLAY SHOWS 
Press and hold the LB/KG and ZERO buttons at the same time, then press the ON/OFF 
button to turn on the scale.  “CAL” followed by a number indicates you can calibrate. 
“CAL” then 
 “XXXXX” 
Press the ZERO button.  “C-100” (for lb mode) or “C-40” (for kg mode) will flash on the 
display and then the display will return to “0”.  
“C-100” or “C-40” 
Place 100 lbs or 40 kg on the scale and when calibration number stabilizes, press the ZERO 
button. “C-200” (for lb mode) or “C-80” (for kg mode) will flash on the display and then the 
display will return to “0”. 
 “C-200” or “C-80” 
Place 200 lbs or 80 kg on the scale and when calibration number stabilizes, press the ZERO 
button. “C-300”  or “C-120” will appear on the display and then the display will return to “0”. 
“C-300” or “C-120” 
Place 300 lbs or 120 kg on the scale and when calibration number stabilizes, press the ZERO 
button. “C-500”  or “C-200” will appear on the display and then the display will return to “0”. 
“C-500” or “C-200” 
Place 500 lbs or 200 kg on the scale and when calibration number stabilizes, press the ZERO 
button. “ END ” will appear on the display and then the scale will automatically turn off. 
“END”