Teledyne MX300-I Manuale Utente

Pagina di 48
Portable Oxygen Monitor  
Operation 
 
 
 Teledyne 
Analytical 
Instruments 
15 
Operation 
Note:  Upon receipt, INSPECT THE ENTIRE UNIT FOR 
DAMAGE. Check the unit and all included accessories for 
broken or loose parts. If damaged, DO NOT USE. Notify 
the shipper, and consult 
Teledyne Analytical Instruments. 
Note:  This equipment is internally powered using 3 AA batteries. 
CAUTION:    THE MX300-I, OXYGEN SENSOR AND ASSOCIATED 
HARDWARE ARE NON-STERILE DEVICES. DO NOT 
AUTOCLAVE THE INSTRUMENT OR SENSOR, AS 
THIS WILL DAMAGE THE EQUIPMENT.  
2.1 Setup 
The MX300-I Portable Oxygen Monitor is suitable for use in many 
medical applications. The unit is equipped with a stand and can be used 
on a tabletop or wall mounted using mounting bracket. An optional pole 
mount clamp is also available. 
To set up and use your MX300-I monitor: 
1.  Install the sensor. 
2.  Install the batteries. 
3.  Calibrate the unit. 
4.  Set the alarms. 
The control keys are designed for easy operation. A LOCK/UNLOCK 
key has been supplied to prevent accidental changes to critical settings. This 
eliminates unwanted changes in calibration or alarm settings from accidental 
touching or bumping of the keys. To further reduce the possibly of incorrect 
adjustments at least two keys must be pressed in order to modify a critical 
calibration or alarm set point value.  
Note:  The ALARM SILENCE () and BATT TEST key continue 
to operate normally when the lock feature is activated.