MAINTENANCE
Each OSP  portable infrared thermometer has been factory tested and calibrated before shipment.
The calibration should be verified and re-adjusted  if the instrument shows an error exceeding the declared specifications
or when a critical active or passive component is replaced (either at the component level or at the board level).
OMEGA 
 engineers will give prompt support for any request of assistance.
16.1
Faulty operating conditions
During the start-up and measuring modes, faulty conditions of the instrument will be announced with coded messages.
If the faulty condition is critical for the application, it is recommended to re-start the pertinent set-up procedure.
All errors which cannot be recovered without the user's knowledge, result in some system action to inform the operator
via message and where possible to restore the system.
Restarting the instrument from -Power ON- may clear the error, but generally such messages are caused by hardware or
software faults, which require the user's action.
Display
Error Type
Solution
Err 1
Internal EEPROM memory error
Contact technical support
Err 2
IR sensor reading error
Verify ambient temperature
(from -5 to +55°C)
Err 3
Calibration data error
Recalibrate unit
OTC
No sensor connected at Tc input
Connect sensor
RJer
Cold-junction sensor reading error
Verify ambient temperature
(from -5 to +55°C)
16.2
Storage
If the instrument has been left unused for a long time, it is recommended to remove the batteries.
Store the instrument  in the original package, at a temperature from -30°C to +60°C, with R.H. less than 90%.
If the instrument has been unused for a month check the battery voltage, and charge the Ni-MH batteries for at least 12
hours with the unit switched off.