Using MediCal Pro
74
6 The black luminance and white luminance measurements may 
influence each other, so that the value of the first step changes when 
the second step is done. Therefore, you have to repeat both steps 
(steps 3 and 4) until both values remain stable.
7 Click  on  Yes to repeat the measurements. Click on No if the values 
remain stable and you wish to proceed with the calibration.
8 For sensors that have an ambient light shielding only: If you selected 
Measure on next calibration in the Preset and the Ambient Light 
Compensation option is checked, the software asks you now to 
remove the sensor from the screen and put it back without Ambient 
Light Shield.
When ready, click on OK.
9 After a few measurements, the software asks to remove the sensor 
once more, and put it back with the Ambient Light Shield.
When ready, click on OK.
10 When the calibration is finished successfully, it displays the following 
message:
11 Click on OK to close the window.
3.5 Setting up a Q/A task schedule