Teledyne MODEL T703 Manuale Utente

Pagina di 287
Specifications and Approvals 
Teledyne API T703 Calibrator Operation Manual 
22    
 
Table 2-2:  T703 Specifications for Ozone Generator 
Flow Rate 
1 to 5 LPM adjustable 
Concentration 
Minimum: 50 ppb at 2 LPM 
Maximum: 5 ppm at 1 LPM 
Output
 
Maximum: 5 ppm LPM 
Minimum: 100 ppb LPM
 
Response Time:
 
180 Sec. (98%)
 
Optical Feedback 
Standard 
 
Table 2-3:  T703 Specifications for O
3
 Photometer 
Ranges 
100 ppb to 10 ppm, user selectable 
Zero Noise 
0.3ppb (RMS) 
Span Noise 
<0.5%  
Lower Detectable Limits 
0.6 ppb (RMS) 
Precision 1.0 
ppb 
Linearity 
1.0% of Full Scale 
Lag Time 
<10 seconds 
Rise/Fall Time 
<20 sec (photometer response) 
Zero Drift 
<1.0 ppb / 7 days 
Span Drift 
<1% / 24 hours; <2% / 7 days 
Flow Rate 
800 cc
3
/min 
+/- 10%
 
 
2.2. CE MARK COMPLIANCE 
EMISSIONS COMPLIANCE 
The Teledyne API’s T703 Photometric O
3
 Calibrator is designed to be fully compliant with: 
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class A, ICES-003 Class A (ANSI 
C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97) Class A. 
Test status: Pending. 
SAFETY COMPLIANCE  
The Teledyne API’s T703 Photometric O
3
 Calibrator is designed to be fully compliant with: 
IEC 61010-1:90 + A1:92 + A2:95, 
Test status: Pending. 
 
 