SETUP Menu 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
102 
5.6. SETUP – CLK: SETTING THE INTERNAL TIME-OF-DAY 
CLOCK 
The T100 has a built-in clock for the AutoCal timer, Time TEST functions, and time 
stamps on COM port messages and DAS data entries. To set the time-of-day, press: 
 
 
 
SETUP X.X 
TIME-OF-DAY CLOCK
 
 TIME  DATE
 
 
 
      
EXIT
 
 
 
 
SETUP X.X
 
      PRIMARY SETUP MENU
 
 
 
 CFG  DAS
  
RNGE
  
PASS  CLK  MORE  
EXIT
 
EXIT
 returns
to the main 
SAMPLE
 display 
Enter Current  
Date-of-Year 
 
 
SETUP X.X
DATE: 01-JAN-02
 
 
 0   1    JAN    0   2
   
 
     ENTR EXIT
 
 
 
 
SETUP X.X
DATE:
 01-JAN-02
 
 
 
0   1    JAN    0   2   
 
     
ENTR
 EXIT
 
 
 
 
SETUP X.X 
TIME-OF-DAY CLOCK 
 
 TIME  DATE 
 
 
      
EXIT 
Enter Current  
Time-of-Day 
 
 
 
SETUP X.X3 
TIME:
 12:00
 
 
 
1   2   : 0   0 
 
 
     
ENTR
 EXIT
 
 
 
 
SETUP X.X 
TIME: 12:00
 
 
 1   2   : 0   0
 
   
 
 
 
 
ENTR 
EXIT
 
 
 
 
SETUP X.X       
PRIMARY SETUP MENU
  
 
 CFG  DAS
  
RNGE
  
PASS  
CLK
  MORE  
EXIT
 
SAMPLE          
ENTER SETUP PASS : 818
 
        
 8        1        8 
 
                      
ENTR
  
 EXIT
 
 SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X   
 
 < TST  TST >  CAL  
 
 
SETUP
 
 
Figure 5-11: 
SETUP – Clock 
In order to compensate for CPU clocks, which may run fast or slow, there is a variable 
to speed up or slow down the clock by a fixed amount every day. To change this 
variable, press: 
06807C DCN6650