Table 6-7:  Default Hessen Status Bit Assignments 
 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
It is possible to assign more than one flag to the same Hessen status bit.  
This allows the grouping of similar flags, such as all temperature 
warnings, under the same status bit. Be careful not to assign conflicting 
flags to the same bit as each status bit will be triggered if any of the 
assigned flags are active.  
STATUS FLAG NAME 
DEFAULT BIT ASSIGNMENT 
WARNING FLAGS 
SAMPLE FLOW WARNING 
0001 
PMT DET WARNING 
0002 
UV LAMP WARNING 
0002 
HVPS WARNING 
0004 
DARK CAL WARNING 
0008 
RCELL TEMP WARNING 
0010 
IZS TEMP WARNING 
0020 
PMT TEMP WARNING 
0040 
INVALID CONC 
0080 
OPERATIONAL FLAGS
 
In Manual Calibration Mode 
0200 
In Zero Calibration Mode 
0400 
In Span Calibration Mode 
0800 
UNITS OF MEASURE FLAGS
 
UGM 0000 
MGM 2000 
PPB 4000 
PPM 6000 
SPARE/UNUSED BITS 
100. 8000 
UNASSIGNED FLAGS
 
Box Temp Warning 
MP Calibration  
Sample Press Warning 
Analog Cal Warning 
System Reset 
Cannot Dyn Zero 
Rear Board Not Detected 
Cannot Dyn Span 
Relay Board Warning 
Instrument Off 