Teledyne API - T100 UV Fluorescence SO2 Analyzer
Communications Setup and Operation
151
Table 6-7: Default Hessen Status Bit Assignments
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
It is possible to assign more than one flag to the same Hessen status bit.
This allows the grouping of similar flags, such as all temperature
warnings, under the same status bit. Be careful not to assign conflicting
flags to the same bit as each status bit will be triggered if any of the
assigned flags are active.
This allows the grouping of similar flags, such as all temperature
warnings, under the same status bit. Be careful not to assign conflicting
flags to the same bit as each status bit will be triggered if any of the
assigned flags are active.
STATUS FLAG NAME
DEFAULT BIT ASSIGNMENT
WARNING FLAGS
SAMPLE FLOW WARNING
0001
PMT DET WARNING
0002
UV LAMP WARNING
0002
HVPS WARNING
0004
DARK CAL WARNING
0008
RCELL TEMP WARNING
0010
IZS TEMP WARNING
0020
PMT TEMP WARNING
0040
INVALID CONC
0080
OPERATIONAL FLAGS
In Manual Calibration Mode
0200
In Zero Calibration Mode
0400
In Span Calibration Mode
0800
UNITS OF MEASURE FLAGS
UGM 0000
MGM 2000
PPB 4000
PPM 6000
SPARE/UNUSED BITS
100. 8000
UNASSIGNED FLAGS
Box Temp Warning
MP Calibration
Sample Press Warning
Analog Cal Warning
System Reset
Cannot Dyn Zero
Rear Board Not Detected
Cannot Dyn Span
Relay Board Warning
Instrument Off