Data Acquisition System (DAS) and APICOM 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
166 
To define the REPORT PERIOD, follow the instruction shown in Figure 7-3, then 
press: 
 
 
Edit Data Channel Menu 
 
 
 
SETUP X.X     
 0) CONC:     ATIMER,     2,        4032, R
 
 PREV  NEXT            INS    DEL
   EDIT    
 PRNT 
   EXIT 
Exits to the main 
Data Acquisition
 
menu. 
 
 
 
SETUP X.X      
NAME:CONC
 
    <SET  
  SET> 
 EDIT
 
   PRINT 
 
EXIT
 
Press SET>  until you reach REPORT PERIOD … 
 
 
 
SETUP X.X     
REPORT PERIODD:DAYS:0
 
    0          0           0   
 
    ENTR    EXIT
 
ENTR
 accepts the new string and 
returns to the previous menu. 
EXIT
 ignores the new string and 
returns to the previous menu. 
Press buttons to set amount of 
time between reports, in hours 
(HH) and/or minutes (MM) 
(max: 23:59). 01:00 sets a 
report to be made every hour. 
Set  the number of days 
between reports (0-365).
 
IIf at any time an invalidl entry is selected (e.g., days > 366) 
the ENTR button will disappear from the display. 
 
 
 
SETUP X.X     
REPORT PERIODD:TIME:01:00
 
    0          1           0          0 
 
    ENTR    EXIT
 
 
 
 
SETUP X.X      
REPORT PERIOD:000:00:05
 
    <SET    SET>
  EDIT 
   PRINT 
 
EXIT
 
Use PREV and NEXT to 
scroll to the data channel 
to be edited. 
 
 
Figure 7-8: DAS – Define the Report Period 
7.2.6. NUMBER OF RECORDS 
Although the DAS is capable of capturing several months worth of data,, the actual 
number of records is also limited by the total number of parameters and channels and 
other settings in the DAS configuration. Every additional data channel, parameter, 
number of samples setting etc. will reduce the maximum amount of data points 
somewhat. In general, however, the maximum data capacity is divided amongst all 
channels (max: 20) and parameters (max: 50 per channel). 
The DAS will check the amount of available data space and prevent the user from 
specifying too many records at any given point. If, for example, the DAS memory space 
can accommodate 375 more data records, the ENTR button will disappear when trying 
to specify more than that number of records. This check for memory space may also 
make an upload of a DAS configuration with APICOM or a Terminal program fail, if 
the combined number of records would be exceeded. In this case, it is suggested to 
either try from the front panel what the maximum number of records can be or use trial-
and-error in designing the DAS script or calculate the number of records using the DAS 
or APICOM manuals. To set the number of records for one channel from the front panel, 
press SETUP-DAS-EDIT-ENTR and the following control button sequence. 