Teledyne T100 Manuale Utente

Pagina di 422
Data Acquisition System (DAS) and APICOM 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
172 
7.4. REMOTE DAS CONFIGURATION VIA APICOM 
Editing channels, parameters and triggering events as described in this section is 
performed via the APICOM remote control program using the graphic interface similar 
to the example shown in Figure 7-14. Refer to Section 8 for details on remote access to 
the T100 analyzer. 
 
Figure 7-14: 
Sample APICOM User Interface for Configuring the DAS 
Once a DAS configuration is edited (which can be done offline and without interrupting 
DAS data collection), it is conveniently uploaded to the instrument and can be stored on 
a computer for later review, alteration or documentation and archival. Refer to the 
APICOM manual for details on these procedures. The APICOM user manual (Teledyne 
API’s P/N 039450000) is included in the APICOM installation file, which can be 
downloaded at http://www.teledyne-api.com/software/apicom/. 
Although Teledyne API recommends the use of APICOM, the DAS can also be 
accessed and configured through a terminal emulation program such as HyperTerminal 
(refer to Figure 7-15). However, all configuration commands must be created following 
a strict syntax or be pasted in from of a text file, which was edited offline and then 
uploaded through a specific transfer procedure. 
06807C DCN6650