Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Remote Operation of the Analyzer 
 
 
183 
To initialize the modem, access the COMMUNICATIONS Menu (refer to Figure 5-13), 
then press: 
 
 
 
 
 
SETUP X.X        
 COM1 MODEM INIT:AT  Y
 &D &H 
 
<SET  
SET>   
 EDIT 
  
 
 
 
  EXIT
 
SETUP X.X         
 COM1 INITIALIZE MODEM 
 
<SET  
SET>   INIT  
  
 
EXIT
 
 
 
 
SETUP X.X 
COM1 MODE:0
 
      
 SET>   
 EDIT
   
       
 
 
  EXIT
 
SETUP X.X          COM1 BAUD RATE:19200
 
 
<SET  
SET>   
 EDIT
 
  
 
   
 
EXIT
 
EXIT
 returns to the 
Communications Menu. 
SETUP X.X              
INITIALIZING MODEM 
 
<SET  
SET>   INIT  
  
 
                  
EXIT
 
 
 
 
SETUP X.X 
COMMUNICATIONS MENU
 
  ID    INET     
 COM1    
 COM2
   
 
      
EXIT
 
Select which 
COM Port is 
tested 
 
 
Figure 8-5: COMM – Initialize the Modem 
 
8.4. COM PORT PASSWORD SECURITY 
In order to provide security for remote access of the T100, a LOGON feature can be 
enabled to require a password before the instrument will accept commands. This is done 
by turning on the SECURITY MODE (refer to Section 5.5). Once the SECURITY 
MODE
 is enabled, the following items apply. 
  A password is required before the port will respond or pass on commands. 
  If the port is inactive for one hour, it will automatically logoff, which can also be 
achieved with the LOGOFF command. 
  Three unsuccessful attempts to log on with an incorrect password will cause 
subsequent logins to be disabled for 1 hour, even if the correct password is used. 
  If not logged on, the only active command is the '?' request for the help screen. 
  The following messages will be returned at logon: 