Teledyne T100 Manuale Utente

Pagina di 422
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Calibration Procedures 
 
 
191 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
If the ZERO or SPAN buttons are not displayed during zero or span 
calibration, the measured concentration value is too different from the 
expected value and the analyzer does not allow zeroing or spanning the 
instrument. Refer to Section 12.4 for more
 information on calibration 
problems. 
9.3. MANUAL CALIBRATION CHECKS 
Informal calibration checks will only evaluate the analyzer’s response curve, but do not 
alter it. It is recommended as a regular maintenance item, to perform calibration checks 
in order to monitor the analyzer’s performance. To carry out a calibration check rather 
than a full calibration, perform the following procedures: 
STEP ONE: Connect the sources of zero air and span gas as shown in Figure 9-1. 
STEP TWO: Perform the zero/span calibration check procedure: 
 
SAMPLE               
STABIL=XXX.X PPB
    
SO2=XXX.X 
 
 
 < TST  TST > 
 CAL 
 
 
 
SETUP
 
ACTION:
Record the SO
2
 
concentration 
reading. 
ACTION:
Supply span gas to the instrument 
SAMPLE              
STABIL=XXX.X PPB
     
SO2=XXX.X  
 
 < TST  TST > 
 CAL 
 
 
 
SETUP
 
ACTION:
Record the SO
2
 
concentration 
reading. 
Wait until 
STABIL
 is 
below 0.5 ppb. 
This may take 
several minutes. 
The value of 
STABIL
 may jump 
significantly. 
Wait until it falls 
below 0.5 ppb. This 
may take several 
minutes. 
SAMPLE              
STABIL=XXX.X PPB
        SO2=XXX.X 
 
 < TST  TST >
 
 CAL 
 
 
 
SETUP
 
The SPAN button appears during the transition from zero to 
span. You may see both the SPAN and the ZERO buttons. 
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X 
 
 < TST  TST >
 
 CAL 
 
 
 
SETUP
 
Scroll the display to the 
STABIL
 test function. 
ACTION:
Supply the instrument with zero gas. 
 
Figure 9-4: Setup for Manual Calibration Checks 
06807C DCN6650