Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Calibration Procedures 
 
 
195 
9.5. MANUAL CALIBRATION WITH IZS OPTION 
Under the best conditions, the accuracy off the SO
2
 effusion rate of the IZS option’s 
permeation tube is about ±5%. This can be subject to significant amounts of drift as the 
tube ages and the amount of SO
2
 contained in the tube is depleted.  Whereas this may be 
sufficient for informal calibration checks, it is not adequate for formal calibrations and is 
not approved for use by the US EPA as a calibration source.  
Therefore, for formal calibrations of an instrument with an IZS option installed the 
following provisions must be followed. 
  Zero air and span gas must be supplied to the analyzer through the sample gas inlet 
  The calibration procedure must be initiated using the CAL button, not the CALZ or 
CALS buttons, using the procedure defined in Section 9.2. 
  Using the CAL button does not activate the zero/span or sample/cal valves of the 
IZS option, thus allowing the introduction of zero air and sample gas through the 
sample port from more accurate, external sources such as a calibrated bottle of SO
2
 
or a Model T700 Dilution Calibrator. 
SAMPLE           RANGE = 500.000 PPB          SO2 =XXX.X   
 
< TST  TST >   
CAL     CALZ     CALS  
        
SETUP
 
Use for formal 
calibration 
operations. 
Use only for 
informal calibration 
checks. 
 
Figure 9-8: Manual Calibration with IZS Option 
 
9.6. MANUAL CALIBRATION CHECKS WITH IZS OR ZERO/SPAN 
VALVES 
Zero and span checks using the zero/span valve or IZS option are similar to that 
described in Section 9.3, with the following exceptions: 
  On units with an IZS option installed, zero air and span gas are supplied to the 
analyzer through the zero gas inlet and from ambient air. 
  On units with a zero/span valve option installed, zero air and span gas are supplied 
to the analyzer through the zero gas and span gas inlets from two different sources. 
  The zero and calibration operations are initiated directly and independently with 
dedicated buttons CALZ and CALS. 