Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Calibration Procedures 
 
 
207 
 
9.10.2. CO
2
 SENSOR CALIBRATION 
Calibration begins with connecting the zero and span gases, then setting the 
concentration values. 
9.10.2.1.  CO
2
 CALIBRATION SETUP 
Bottled gases are connected as follows: 
at
 Sp
an 
Conce
ntration
at 
Spa
Concen
tration
Source of
(Remove during 
calibration)
3-way
Valve
Manual 
Control Valve
Chassis
NO
NC
COM
 
Figure 9-17: 
CO
2
 Sensor Calibration Set Up 
CO
2
 SENSOR ZERO GAS: Teledyne API recommends using pure N
2
 when calibration 
the zero point of your CO
2
 sensor option. 
CO
2
  SENSOR SPAN GAS: Teledyne API recommends using 16% CO
2
 in N
2
 when 
calibration the span point of your CO
2
 sensor is 20%. 