Teledyne T100 Manuale Utente

Pagina di 422
Teledyne API
 - T100 UV Fluorescence SO2 Analyzer 
EPA Protocol Calibration 
 
 
215 
10.1.7.  SUMMARY OF QUALITY ASSURANCE CHECKS 
The following items should be checked on a regularly scheduled basis to assure high 
quality data from the T100. Refer to Table 10-3 for a summary of activities. Also the 
QA Handbook should be checked for specific procedures. 
Table 10-3: 
Activity Matrix for Quality Assurance Checks 
CHARACTERISTIC 
ACCEPTANCE LIMITS 
FREQUENCY AND METHOD OF 
MEASUREMENT 
ACTION IF REQUIREMENTS ARE 
NOT MET 
Shelter temperature  Mean temperature between 
22
o
C and 28
o
C (72
o
 and 
82
o
F), daily fluctuations not 
greater than ±2
o
Check thermograph chart 
weekly for variations greater 
than ±2
o
C (4
o
F) 
Mark strip chart for the affected 
time period 
 Repair or adjust temperature 
control 
Sample introduction 
system 
No moisture, foreign 
material, leaks, obstructions; 
sample line connected to 
manifold 
Weekly visual inspection 
Clean, repair, or replace as 
needed 
Recorder 
Adequate ink & paper  
Legible ink traces 
Correct chart speed and 
range  
Correct time 
Weekly visual inspection 
Replenish ink and paper supply 
Adjust time to agree with clock; 
note on chart 
Analyzer operational 
settings 
TEST measurements at 
nominal values 
2. T100 in SAMPLE mode 
Weekly visual inspection  
Adjust or repair as needed 
Analyzer operational 
check 
Zero and span within 
tolerance limits as described 
in Subsec. 9.1.3 of Sec. 
2.0.9 (Q.A. Handbook) 
Level 1 zero/span every 2 
weeks; Level 2 between Level 
1 checks at frequency desired 
analyzer by user 
Find source of error and repair 
After corrective action, re-
calibrate analyzer 
Precision check 
Assess precision as 
described in Sec. 2.0.8 and 
Subsec. 3.4.3 (Ibid.) 
Every 2 weeks, Subsec. 3.4.3 
(Ibid.) 
Calc, report precision, Sec. 2.0.8 
(Ibid.) 
10.2. LEVEL 1 CALIBRATIONS VERSUS LEVEL 2 CHECKS 
Essential to quality assurance are scheduled checks for verifying the operational status 
of the monitoring system. The operator should visit the site at least once each week. It is 
recommended Level 1 zero and span check conducted on the analyzer every two weeks. 
Level 2 zero and span checks should be conducted at a frequency desired by the user. 
Definitions of these terms are given in Table 10-4. 
In addition, an independent precision check between 0.08 and 0.10 ppm must be carried 
out at least once every two weeks. Table 10-3 summarizes the quality assurance 
activities for routine operations. A discussion of each activity appears in the following 
sections.  
To provide for documentation and accountability of activities, a checklist should be 
compiled and then filled out by the field operator as each activity is completed. 
06807C DCN6650