219 
10.5. DYNAMIC MULTIPOINT SPAN CALIBRATION 
Dynamic calibration involves introducing gas samples of known concentrations to an 
instrument in order to record the instruments performance at a predetermined sensitivity 
and to derive a calibration relationship. A minimum of three reference points and one 
zero point uniformly spaced covering 0 to 90 percent of the operating range are 
recommended to define this relationship. 
The analyzer's recorded response is compared with the known concentration to derive 
the calibration relationship.  
To perform a precision check, the instrument set up, sources of zero air and sample gas 
should conform to those described in Section 9.2. 
Follow the procedures described in Section 9.2 for calibrating the zero points.   
For each mid point: 
Wait until 
STABIL
 falls 
below 0.5 ppb. 
This may take 
several minutes. 
SAMPLE              
STABIL=X.XXX PPB
          SO2 =XXX.X  
 
 < TST  TST >  CAL
 
 CALZ   
CALS
 
 
SETUP
 
Record the SO
2
 
reading as 
displayed on the 
instrument’s front 
panel 
ACTION: 
Allow calibration gas diluted to proper concentration for 
Midpoint N to enter the sample port
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X  
 
 < TST  
TST
 >  
CAL
 
 
 
 
SETUP
 
 SAMPLE             
STABIL=X.XXX PPB
          SO2 =XXX.X 
 
 < TST  TST >  
CAL
 
 
 
 
SETUP
 
Set the Display to show the 
STABIL
 test function. 
This function calculates the 
stability of the SO
2
  
measurement 
SPAN CAL M
         RANGE = 500.0 PPB        
 SO
2
 XXX.X 
 
 < TST  TST > 
 ZERO
 
 SPAN
   CONC
   
EXIT 
ACTION: 
Allow Calibration Gas diluted to 
proper concentration for 
Midpoint N+1 to enter the sample 
port 
Press EXIT to 
Return to the 
Main SAMPLE 
Display 
 
Figure 10-1: 
Dynamic Multipoint Span Calibration 
06807C DCN6650