Teledyne T100 Manuale Utente

Pagina di 422
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Instrument Maintenance 
 
 
233 
11.3.7. PERFORMING A SAMPLE FLOW CHECK 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
Use a separate, calibrated flow meter capable of measuring flows 
between 0 and 1000 cm³/min to measure the gas flow rate though the 
analyzer. For this procedure, do not refer to the built in flow 
measurement shown in the front panel display screen.  
Sample flow checks are useful for monitoring the actual flow of the instrument, to 
monitor drift of the internal flow measurement. A decreasing, actual sample flow may 
point to slowly clogging pneumatic paths, most likely critical flow orifices or sintered 
filters. To perform a sample flow check:  
1.  Disconnect the sample inlet tubing from the rear panel SAMPLE port (Figure 3-4)
2.  Attach the outlet port of a flow meter to the sample inlet port on the rear panel. 
Ensure that the inlet to the flow meter is at atmospheric pressure. 
3.  The sample flow measured with the external flow meter should be 650 cm³/min 
 
10%.  
4.  Low flows indicate blockage somewhere in the pneumatic pathway. Refer to 
11.3.8. HYDROCARBON SCRUBBER (
KICKER
There are two possible types of problems that can occur with the scrubber: pneumatic 
leaks and contamination that ruins the inner tube’s ability to absorb hydrocarbons. 
11.3.8.1. CHECKING THE SCRUBBER FOR LEAKS 
Leaks in the outer tubing of the scrubber can be found using the procedure described in 
Section  11.3.6.  Use the following method to determine if a leak exists in the inner 
tubing of the scrubber. 
This procedure requires a pressurized source of air (chemical composition is 
unimportant) capable of supplying up to 15 psiA and a leak checking fixture such as the 
one illustrated in Figure 11-3. 
 
Vacuum/Pressure 
Gauge 
Manual Shut-Off 
Valve
Needle Valve 
FROM PUMP or 
PRESSURIZED 
AIR SOURCE 
TO SCRUBBER 
 
Figure 11-3: 
Simple Leak Check Fixture 
06807C DCN6650