Teledyne T100 Manuale Utente

Pagina di 422
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Troubleshooting & Service 
 
 
239 
Warning Message 
Fault Condition 
Possible Causes 
SAMPLE PRES WARN 
Sample Pressure is <10 in-
Hg or 
 > 35 in-Hg
1
 
If sample pressure is < 10 in-hg: 
o
  Blocked particulate filter 
o
  Blocked sample inlet/gas line 
o
  Failed pressure sensor/circuitry 
If sample pressure is > 35 in-hg: 
o
  Blocked vent line on pressurized sample/zero/span gas supply 
o
  Bad pressure sensor/circuitry 
SYSTEM RESET 
The computer has 
rebooted. 
 
This message occurs at power on.   
If it is confirmed that power has not been interrupted: 
Failed +5 VDC power,  
Fatal error caused software to restart   
Loose connector/wiring  
UV LAMP WARNING 
The UV lamp intensity is < 
600mV or > 4995 mV 
UV lamp is bad 
Reference detector is bad or out of adjustment. 
Mother board analog sensor input circuitry has failed. 
Fogged or damaged lenses/filters in UV light path 
A/D converter circuitry failure 
Light leak in reaction cell 
Shutter solenoid stuck closed 
Normally 29.92 in-Hg at sea level decreasing at 1 in-Hg per 1000 ft of altitude  
(with no flow – pump disconnected). 
 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
A failure of the analyzer’s CPU, motherboard or power supplies can 
result in any or ALL of the above messages. 
12.1.2. FAULT DIAGNOSIS WITH TEST FUNCTIONS 
Besides being useful as predictive diagnostic tools, the TEST functions, viewable from 
the front panel, can be used to isolate and identify many operational problems when 
combined with a thorough understanding of the analyzer’s principles of operation (refer 
to Section 13. We recommend use of the APICOM remote control program (Section 7) 
to download, graph and archive TEST data for analysis, and long-term monitoring of 
diagnostic data. 
The acceptable ranges for these test functions are listed in Table A-3 in Appendix A-3. 
The actual values for these test functions on checkout at the factory were also listed in 
the  Final Test and Validation Data Sheet, which was shipped with the instrument. 
Values outside the acceptable ranges indicate a failure of one or more of the analyzer’s 
subsystems. Functions with values that are within the acceptable range but have 
significantly changed from the measurements recorded on the factory data sheet may 
also indicate a failure or a maintenance item. 
A problem report worksheet has been provided in Appendix C to assist in recording the 
value of these test functions. Table 12-2 contains some of the more common causes for 
these values to be out of range. 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
A value of “XXXX” displayed for any of these TEST functions indicates an 
OUT OF RANGE reading. 
 
06807C DCN6650