Troubleshooting & Service 
 
 
243 
12.2. STATUS LEDS 
Several color-coded, light-emitting diodes (LEDs) are located inside the instrument to 
determine if the analyzer’s CPU, I
2
C communications bus and relay board are 
functioning properly. 
12.2.1. MOTHERBOARD STATUS INDICATOR (WATCHDOG) 
DS5, a red LED on the upper portion of the motherboard, just to the right of the CPU 
board, flashes when the CPU is running the main program. After power-up, DS5 should 
flash on and off about once per second. If characters are written to the front panel 
display but DS5 does not flash then the program files have become corrupted. Contact 
Teledyne API’s Technical Support department. 
If DS5 is not flashing 30 - 60 seconds after a restart and no characters have been written 
to the front panel display, the firmware may be corrupted or the CPU may be defective. 
If DS5 is permanently off or permanently on, the CPU board is likely locked up and 
should the analyzer not respond (either with locked-up or dark front panel), then replace 
the CPU. 
 
Motherboard
 
CPU Status LED 
 
Figure 12-3: 
CPU Status Indicator 
12.2.2. CPU STATUS INDICATORS 
The LEDs on the CPU card (Figure 13-14) are described as follows: 
Power LED 
Red 
normally lit 
IDE LED 
Green 
lit when active (read or write) 