Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Troubleshooting & Service 
 
 
251 
Table 12-5: DC Power Supply Acceptable Levels 
CHECK RELAY BOARD TEST POINTS 
FROM TEST POINT
TO TEST POINT 
POWER 
SUPPLY 
VOLTAGE
NAME #  NAME  # 
MIN V 
MAX V 
PS1 +5  DGND 
1  +5  2 
+4.80 
+5.25 
PS1 +15  AGND 3  +15  4 
+13.5 
+16.0 
PS1 -15  AGND 3  -15V 5 
-14.0 
-16.0 
PS1 AGND  AGND  3  DGND  1 -0.05  +0.05 
PS1 Chassis DGND 1 Chassis N/A 
-0.05 +0.05 
PS2 +12 +12V 
Ret 
6  +12V  7 
+11.8 
+12.5 
PS2 DGND +12V 
Ret 6  DGND  1 -0.05  +0.05 
 
12.6.3. I
2
C BUS 
Operation of the I
2
C bus can be verified by observing the behavior of D1 on the relay 
PCA & D2 on the Valve Driver PCA . Assuming that the DC power supplies are 
operating properly, the I
2
C bus is operating properly if: D1 on the relay PCA and D2 of 
the Valve Driver PCA are flashing 
There is a problem with the I
2
C bus if both D1 on the relay PCA and D2 of the Valve 
Driver PCA are ON/OFF constantly. 
12.6.4. TOUCH-SCREEN INTERFACE 
Verify the functioning of the touch screen by observing the display when pressing a 
touch-screen control button. Assuming that there are no wiring problems and that the 
DC power supplies are operating properly, but pressing a control button on the touch 
screen does not change the display, any of the following may be the problem: 
  The touch-screen controller may be malfunctioning.  
   The internal USB bus may be malfunctioning.  
You can verify this failure by logging on to the instrument using APICOM or a terminal 
program. If the analyzer responds to remote commands and the display changes 
accordingly, the touch-screen interface may be faulty. 
12.6.5. LCD DISPLAY MODULE 
Verify the functioning of the front panel display by observing it when power is applied 
to the instrument. Assuming that there are no wiring problems and that the DC power 
supplies are operating properly, the display screen should light and show the splash 
screen and other indications of its state as the CPU goes through its initialization 
process. 
 