Teledyne T100 Manuale Utente

Pagina di 422
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Troubleshooting & Service 
 
 
255 
12.6.9.2. MODEM OR TERMINAL OPERATION 
These are the general steps for troubleshooting problems with a modem connected to a 
Teledyne API analyzer. 
  Check cables for proper connection to the modem, terminal or computer. 
  Check the correct position of the DTE/DCE switch as described in Section 6.1. 
  Check the correct setup command (refer to Section 8.3). 
  Verify that the Ready to Send (RTS) signal is at logic high. The T100 sets Pin 7 
(RTS) to greater than 3 volts to enable modem transmission. 
  Ensure that the baud rate, word length, and stop bit settings between modem and 
analyzer match (refer to Sections 6.2.2 and 8.3). 
  Use the RS-232 test function to send “w” characters to the modem, terminal or 
computer. Refer to Section 6.2.3. 
  Get your terminal, modem or computer to transmit data to the analyzer (holding 
down the space bar is one way). The green LED on the rear panel should flicker as 
the instrument is receiving data. 
  Ensure that the communications software is functioning properly. 
Further help with serial communications is available in a separate manual “RS-232 
Manual”, Teledyne API’s P/N 013500000, available online at 
http://www.Teledyne-
api.com/manuals/
12.6.10. SHUTTER SYSTEM   
To check the functionality of the UV light Shutter by manually activating it: 
 
EXIT 
4x’s  to return 
 to the  
SAMPLE 
display 
 
 
 
DIAG  I / O        
36) UVLAMP_SIGNAL= 3.4 MV
 
 
  PREV NEXT
 
JUMP
 
                        
PRNT 
EXIT
 
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X     
 
 < TST  TST >  CAL  
 
 
SETUP
 
SAMPLE          
ENTER SETUP PASS : 818
 
 
        
 8        1        8 
 
                       
ENTR
  
 EXIT
 
SETUP X.X 
 
      
PRIMARY SETUP MENU
 
  
 
 CFG  DAS
  
RNGE
  
PASS  CLK  
MORE
  
   EXIT
 
 
 
 
DIAG  
SIGNAL I / O  
 
 PREV NEXT JUMP
 
 
                       ENTR
 
  
EXIT
 
 
 
 
DIAG  I / O           
0)
 
EXT_ZERO_CAL=OFF
 
 
PREV  NEXT
 JUMP 
                      
PRNT  EXIT
 
Press these buttons 
until 32 is displayed 
 
 
 
DIAG  I / O 
JUMP TO: 01
 
 
 
 0     1                       
 
        ENTR
  EXIT
 
Activate the 
Dark Shutter 
 
 
 
DIAG  I / O        
32) DARK_SHUTTER=OFF
 
 
  PREV NEXT
 
JUMP
 
                OFF   
PRNT 
EXIT
 
SETUP X.X     
SECONDARY SETUP MENU
  
 
 
COMM  VARS  
DIAG  
 
 
      
   EXIT
 
EXIT
 returns 
to the main 
SAMPLE
 display  
 
 
 
DIAG  I / O        
32) DARK_SHUTTER=ON
 
 
  PREV NEXT
 JUMP 
                 
ON
   
PRNT 
EXIT
 
 
 
 
DIAG  I / O 
JUMP TO: 32
 
 
 
 3      2                       
 
        ENTR
  EXIT
 
Press these buttons 
until 36 is displayed 
UV LAMP_SIGNAL
should be  
<20 mV 
 
06807C DCN6650