Pagina di 422
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Troubleshooting & Service 
 
 
277 
QUESTION ANSWER 
How do I program and use them? 
How long does the sample pump last?   The sample pump should last about one year and the pump diaphragms should to 
be replaced annually or when necessary. Use the PRES test function displayed via 
the front panel to see if the diaphragm needs replacement (refer to Section 12.1.2).  
Do I need a strip chart recorder or 
external data logger? 
No, the T100 is equipped with a very powerful internal data acquisition system. 
Section 0 describes the setup and operation in detail. 
 