Teledyne T100 Manuale Utente

Pagina di 422
Principles of Operation 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
284 
 
Light Output 
Portal 
Zinc-Vapor 
Plasma Arc
Vacuum 
Jacket 
Dual Bore 
 
Figure 13-3: Source UV Lamp Construction 
 
13.1.4. THE REFERENCE DETECTOR 
A vacuum diode, UV detector that converts UV light to a DC current is used to measure 
the intensity of the excitation UV source lamp. It is located directly across from the 
source lamp at the back of a narrow tube-shaped light trap, which places it directly in the 
path of the excitation UV light. A window transparent to UV light provides an air-proof 
seal that prevents ambient gas from contaminating the sample chamber. The shape of the 
light trap and the fact that the detector is blind to wavelengths other than UV no extra 
optical filtering is needed.  
13.1.5. THE PMT 
The amount of fluoresced UV produced in the sample chamber is much less than the 
intensity of excitation UV source lamp (refer to Figure 13-4). Therefore a much more 
sensitive device is needed to detect this light with enough resolution to be meaningful. 
The T100 uses a Photo Multiplier Tube or PMT for this purpose.  
A PMT is typically a vacuum tube containing a variety of specially designed electrodes. 
Photons enter the PMT and strike a negatively charged photo cathode causing it to emit 
electrons. These electrons are accelerated by a high voltage applied across a series of 
special electrodes called dynodes that multiply the amount of electrons until a useable 
current signal is generated. This current increases or decreases with the amount of 
detected light (refer to Section 13.5.3 for more details regarding the electronic operation 
of the PMT). 
13.1.6. UV LAMP SHUTTER & PMT OFFSET 
Inherent in the operation of both the reference detector and the PMT are a minor 
electronic offsets.  The degree of offset differs from detector to detector and from PMT 
to PMT and can change over time as these components age.    
To account for these offsets the T100 includes a shutter, located between the UV Lamp 
and the source filter that periodically cuts off the UV light from the sample chamber.  
This happens every 30 minutes. The analyzer records the outputs of both the reference 
detector and the PMT during this dark period and factors them into the SO
2
 
concentration calculation. 
06807C DCN6650