Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Principles of Operation 
 
 
293 
13.4. PNEUMATIC OPERATION 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
It is important that the sample airflow system is leak-tight and not 
pressurized over ambient pressure. Regular leak checks should be 
performed on the analyzer as described in the maintenance schedule, 
Table 11-1. Procedures
 for correctly performing leak checks can be found 
in Section 11.3.6. 
 
I
MPORTANT
 
Relative Pressure versus Absolute Pressure 
In this manual vacuum readings are given in inches of mercury absolute 
pressure (in-Hg-A), i.e. indicate an absolute pressure referenced against 
zero (a perfect vacuum). 
13.4.1. SAMPLE GAS FLOW 
The Flow of gas through the T100 UV Fluorescence SO
2
 Analyzer is created by a small 
internal pump that pulls air though the instrument. 
 
VACUUM
 M
ANI
F
O
L
FLOW 
CONTROL 
ASSY 
EXHAUST TO OUTER
LAYER OF KICKER 
SAMPLE  
gas inlet 
SAMPLE FILTER
Chassis
EXHAUST  
gas outlet
 
KICKER EXHAUST 
TO PUMP
HYDROCARBON 
SCRUBBER 
(KICKER)
 
UV  
LAMP 
PMT 
SAMPLE 
CHAMBER 
FLOW 
SENSOR
FLOW / PRESSURE 
SENSOR PCA 
SAMPLE 
PRESSURE 
SENSOR 
PUMP 
CRITICAL 
FLOW 
ORIFICE
 
 
Figure 13-10:  Gas Flow and Location of Critical Flow Orifice 