Principles of Operation 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
302 
13.5.3. PHOTO MULTIPLIER TUBE (PMT) 
The T100 uses a photo multiplier tube (PMT) to detect the amount of fluorescence 
created by the SO
2
 and O
3
 reaction in the sample chamber. 
PMT Housing End Plate 
This is the entry to the PMT Exchange
PMT Preamp PCA 
High voltage Power Supply 
(HVPS) 
PMT 
PMT Cold Block 
Connector to PMT 
Pre Amp PCA 
12V Power 
Connector 
Cooling Fan 
Housing 
TEC Driver PCA 
PMT Heat Exchange Fins
Light from Reaction 
Chamber shines 
through hole in side 
of  Cold Block 
Insulation Gasket 
PMT Power Supply 
& Aux. Signal 
Connector 
PMT Output 
Connector 
Thermo-Electric Cooler 
(TEC) 
PMT Temperature 
Sensor
O-Test LED 
 
Figure 13-17:  PMT Housing Assembly  
A typical PMT is a vacuum tube containing a variety of specially designed electrodes. 
Photons from the reaction are filtered by an optical high-pass filter, enter the PMT and 
strike a negatively charged photo cathode causing it to emit electrons. A high voltage 
potential across these focusing electrodes directs the electrons toward an array of high 
voltage dynodes. The dynodes in this electron multiplier array are designed so that each 
stage multiplies the number of emitted electrons by emitting multiple, new electrons. 
The greatly increased number of electrons emitted from one end of electron multiplier is 
collected by a positively charged anode at the other end, which creates a useable current 
signal. This current signal is amplified by the preamplifier board and then reported to the 
motherboard. 
06807C DCN6650