Teledyne T100 Manuale Utente

Pagina di 422
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Principles of Operation 
 
 
313 
13.6. FRONT PANEL/DISPLAY INTERFACE 
Users can input data and receive information directly through the front panel touch-
screen display. The LCD display is controlled directly by the CPU board. The 
touchscreen is interfaced to the CPU by means of a touchscreen controller that connects 
to the CPU via the internal USB bus and emulates a computer mouse. 
CPU
LCD Display
and
Touchscreen
TFT BIAS 
Supply
10.4, -7.0, 16, 4V
3.3V
Touch Screen Controller
USB2 HUB
USB & 5V
USB-1
USB-2
Powered
Powered
LVDS 
Receiver
LVDS 
Transmitter 
Board
USB Master
Back-Light 
Supply
18 Bit TTL Data
+5V
BL Cont.
PWM 
  LAN     COM4       USB4
Aux I/O PCA
Ethernet
USB Slave
Front Panel Interface PCA
Utility 
Controller
Lang.
Remote
Local
Ethernet Port
USB 
Type B Port
Analog Input 
Terminal Block
 
Figure 13-23:  Front Panel and Display Interface Block Diagram 
13.6.1. LVDS TRANSMITTER BOARD 
The LVDS (low voltage differential signaling) transmitter board converts the parallel 
display bus to a serialized, low voltage, differential signal bus in order to transmit the 
video signal to the LCD interface PCA. 
13.6.2. FRONT PANEL INTERFACE PCA 
The front panel interface PCA controls the various functions of the display and 
touchscreen. For driving the display it provides connection between the CPU video 
controller and the LCD display module. This PCA also contains: 
  power supply circuitry for the LCD display module  
  a USB hub that is used for communications with the touchscreen controller and the 
two front panel USB device ports 
  the circuitry for powering the display backlight 
06807C DCN6650