Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Specifications, Approvals & Compliance 
 
 
31 
 
2.2. EPA EQUIVALENCY DESIGNATION 
The T100 Analyzer is designated as Reference Method Number EQSA-0495-100 as per 
40 CFR Part 53 when operated under the following conditions: 
  Range: Any range from 50 parts per billion (ppb) to 10 parts per million (ppm) 
  Ambient temperature range of 5
 o
C to 40 
o
  Line voltage range of 100-120 VAC or 220-240 VAC, at 50 or 60 Hz 
  Sample filter: Equipped with PTFE filter element in the internal filter assembly 
  Sample flow of 650 +/- 65 cm
3
/min 
  Vacuum pump (internal) capable of 14"Hg Absolute pressure @ 1 slpm or better 
  Software settings: 
Dynamic span 
OFF 
Dynamic zero 
OFF 
Dilution factor 
OFF 
AutoCal 
ON or OFF 
Dual range 
ON or OFF 
Auto-range 
ON or OFF 
Temp/Pressure compensation ON 
 
Under this designation, the analyzer may be operated with or without the following 
optional equipment: 
  Rack mount with chassis slides 
  Rack mount without slides, ears only 
  Zero/span valve options. 
  Internal zero/span (IZS) option with either: 
  SO
2
 permeation tube - 0.4ppm at 0.7 liter per minute; certified/uncertified, 
or 
  SO
2
 permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified. 
Under the designation, the IZS option cannot be used as the source of 
calibration 
  4-20mA isolated analog outputs 
  Status outputs 
  Control inputs 
  RS-232 output 
  Ethernet output 
  Zero air scrubber 
  4-20mA, isolated output 