Teledyne API - T100 UV Fluorescence SO2 Analyzer  
68 
 
CAUTION - GENERAL SAFETY HAZARD 
Do not look at the UV lamp while the unit is operating. UV light can 
cause eye damage. Always use safety glasses made from UV blocking 
material whenever working with the UV Lamp. (Generic plastic glasses 
are not adequate). 
3.4.1. STARTUP 
After the electrical and pneumatic connections are made, an initial functional check is in 
order. Turn on the instrument.  The pump and exhaust fan should start immediately.  
The display will show a momentary splash screen of the Teledyne API logo and other 
information during the initialization process while the CPU loads the operating system, 
the firmware and the configuration data. 
The analyzer should automatically switch to Sample Mode after completing the boot-up 
sequence and start monitoring the
 
gas. However, there is an approximately one hour 
warm-up period before reliable gas measurements can be taken. During the warm-up 
period, the front panel display may show messages in the Parameters field.  
3.4.2. WARNING MESSAGES 
Because internal temperatures and other conditions may be outside the specified limits 
during the analyzer’s warm-up period, the software will suppress most warning 
conditions for 30 minutes after power up.  If warning messages persist after the 60 
minutes warm up period is over, investigate their cause using the troubleshooting 
guidelines in Section 12.1.1. 
To view and clear warning messages, press:   
 
Figure 3-23: 
Warning Messages 
 
 