Canon WITE32 Manuale Utente

Pagina di 29
 
WITE32 Release Notes Version 3.11 
 
 
2.6 
TAA Calibration Through Programmable Differentiator  
Starting from WITE32 version 3.11, the calibration of amplification of programmable differentiator filters is not 
performed during TAA calibration. All programmable differentiators are assumed to have the same amplification at 
their cutoff frequencies as the Open filter. This means that if the spectral content of the signal is located below the 
cutoff frequency of a programmable differentiator, then the amplitude measured through this differentiator will 
decrease (approximately linearly) with the increase of the cutoff frequency. 
In previous releases of WITE32, the amplification of the programmable differentiator was calibrated during TAA 
calibration regardless of the current cutoff frequency and the calibration signal spectral content. This led to 
unpredictable measurement results in the case of changes of the cutoff frequency or the signal frequency. 
2.7 
Error / Warning Report Modification 
1.  The information about WITE32 software version is added to the standard WITE32 error message box, when 
shown in the More Info mode. The WITE32 software version is printed on the first line of the error report. 
Figure 16 
Error / Warning Dialog Box 
2. The 
existing 
Errors at The Test End option is extended.  In previous versions of WITE32, if this option is 
enabled, the buffered messages are only shown in a separate dialog box at the end of the production test.  In 
WITE32 version 3.11 you can specify a log file name, such that the buffered error messages are also appended to 
the log file. 
Figure 17 
Test Options Dialog Box with Error Log File Name Control  
 
 
 
 
Guzik Technical Enterprises 
20