SECTION 6—MAINTENANCE
23
SECTION 6—MAINTENANCE
 WARNING
The Invacare concentrators are specifically designed to 
minimize routine preventive maintenance. Only qualified 
personnel should perform preventive maintenance on the 
concentrator. 
Power should be disconnected before beginning preventive 
maintenance on the Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be 
performed according to the maintenance record guidelines. In places 
with high dust or soot levels, maintenance may need to be performed 
more often. Refer to Preventive Maintenance Record on page 25.
Routine Maintenance
NOTE: The following routine maintenance should be performed 
more frequently. See recommended intervals in each section and 
preventive maintenance record. 
Cleaning the Cabinet Filter
CAUTION
DO NOT operate the concentrator without the filter installed.
NOTE: For this procedure, refer to FIGURE 6.1 on page 24. 
NOTE: There is one cabinet filter located on the back of the cabinet.
1. Remove the filter and clean at least once a week depending 
on environmental conditions.
NOTE: Environmental conditions that may require more frequent 
cleaning of the filters include but are not limited to: high dust, air 
pollutants, etc.