Operation 
H25-IR 
 
30
 
Instruction 3015-4342 
4.2.3 Range Selection 
The H25-IR has three sensitivity ranges: SM (small), MED (medium), and LG (large). The range used 
when searching for a leak depends on the size of the leak and the type of gas being monitored.  
 
Press the RANGE button until the desired range is displayed on either the instrument’s Data Display 
screen, or on the LCD of the Smart probe. 
 
 
Instrument Display 
Smart Probe Display 
 
 
 
 
 
 
NOTE: 
The range setting has no affect on the accuracy of the gas reading as displayed on the instrument 
when making a leak-rate measurement. It does, however, change the sensitivity of the bar graph at the 
bottom of the Data Display screen, the audible tone frequency as heard from the speaker, and the blink rate 
of the red LED on the hand-held probe. 
 
The “SM” range is the highest sensitivity level, and is most likely used for fluorine based gases like R134a 
and most fluorine based refrigerant blends. 
 
The “MED” range is used for chlorine and chlorine based gases like R12, R22, R123, etc.  This position is 
also useful for locating large R134a or other fluorine based refrigerant leaks of 1−2 Oz/yr. 
 
The “LG” range is used to locate very large leaks.  
 
TIP:
 Blowing out the test area with shop air may enable the operator to locate a large leak more quickly. 
 
 
4.2.4 Setup Selection 
There are four user-defined Setups that allow the operator to quickly switch between frequently used 
combinations of Gas Type, Range, Feedback Mode, Setpoint, and the Auto Zero Off condition (refer to 
When first turned ON, the instrument returns to the last Setup selected. To select a different Setup, first 
press the SETUP button to display the Setup Selection menu. Next, use the Up and Down buttons to 
highlight the desired Setup, and then press ENT to make the selection and be returned to the Data Display 
screen. The name of the newly selected Setup will appear on the Data Display screen. 
 
 
 
 
 
NOTE: 
The names listed in the Setup Selection menu are user defined, and may vary from what is 
shown above. The number of each setup, however, is what appears on the display of the Smart probe. 
 
  
Instrument Display 
Smart Probe Display 
 
 
 
 
 
 
SETUP NO. 1 
 
SETUP NO. 2 
 
SETUP NO. 3 
 
SETUP NO. 4 
Setup Name  
and Number
0.00  SM
SRCH (1)
Oz/yr 
R134A 
SETUP NO 
SM SRCH
Range: SM, MED, LG
0.00  SM
SRCH (1)
Oz/yr 
R134A 
SETUP NO 
SM  SRCH