4.3  What to Do when a Fault Occurs 
When the instrument’s diagnostics detects that a fault has occurred, the word “FAULT” appears in place of 
the Gas Type, and an audible beep occurs every 2 seconds.  
 
 
 
 
 
 
The cause of the fault can be determined by viewing the Diagnostic menu’s Last Fault function. To display 
the Diagnostics menu, first press both the ENT and ESC buttons on either the instrument’s front panel or 
Smart probe at the same time to display the Main menu. Next, use the Up and Down buttons to highlight 
“DIAGNOSTICS” and then press ENT to display the Diagnostics menu.  
 
Main Menu:  
 
SYSTEM 
 LOG 
 SETUPS 
 CAL 
DATA 
 DIAGNOSTICS 
 FACTORY
 
 
 
 
 
Select “LAST FAULT” to display the Last Fault screen. Refer to Section 5.1.2 Last Fault for a listing of the 
type of faults that are tracked by the diagnostics, and for information on how to possibly correct the fault. 
 
Once the cause of the fault has been corrected, the FAULT message and the audible beep are 
automatically turned OFF. 
 
 
4.4  What is PCHK? 
The instrument maintains its accuracy under varying atmospheric conditions by performing an 
atmospheric pressure check either once every hour, or immediately after calibration (refer to Section 
Error! Reference source not found. Error! Reference source not found.
). When a PCHK is 
performed, the unit beeps once, the pump stops, a valve opens exposing the pressure sensor to the 
atmosphere, and “PCHK” appears at the bottom-right of the display. During this time the instrument 
measures the current atmospheric pressure, and then uses this value along with the current temperature 
and the ideal gas law to compute the gas level. When the pressure check is complete (approximately 5 
seconds), the unit beeps, the valve closes, and the pump restarts. 
 
 
 
 
 
 
 
4.5 RS-232 Communication 
The RS-232 connector located on the rear of the instrument (refer to Section 1.7.2 Rear View) is used to 
flash the instrument’s memory. Refer to Section 5.3 Flashing Memory
 
 
 
Oz/yr 
R134A 
SETUP 
          PCHK 
Pressure check being performed
Diagnostics Menu:  
 
SENSOR DATA 
 LAST 
FAULT 
 
IR EMITTER 
 PRESSURE 
Fault has occurred
Oz/yr 
FAULT 
SETUP 
SM  SRCH