Bacharach H25-IR Manuale Utente

Pagina di 60
Maintenance 
H25-IR 
 
40
 
Instruction 3015-4342 
5.2.1 Auto Gain 
 
 
 
 
IMPORTANT!
 The instrument must be sampling fresh air (zero gas) when the Auto Gain function 
is selected. 
Selecting the Auto Gain function automatically sets the IR Detector’s voltage to its baseline 0 ppm level by 
controlling the instrument’s digital potentiometer (DIGIPOT) circuit. 
5.2.2 DIGIPOT 
 
 
 
 
IMPORTANT!
 The instrument must be sampling fresh air (zero gas) when adjusting the DIGIPOT value. 
 
The DIGIPOT function allows manual and a more precise adjustment of the digital potentiometer, which 
controls the voltage of the IR Detector. Use the Up and Down buttons to adjust DIGIPOT value until the 
IR VOLT reading is 4.200 ±0.0500 volts. Press ENT to save the new DIGIPOT value. 
5.2.3 Leak Rate 
 
 
 
 
The Leak Rate value is set to match the leak rate of the internal leak source as indicated on the tag 
attached to the leak source. Use the Left and Right buttons to move the blinking cursor across the screen 
to highlight the digit to be modified, and then use the Up and Down buttons to modify that digit. Press 
ENT to save the new Leak Rate value. 
IMPORTANT!
  Be sure to enter the leak rate using the same units of measure as displayed on the LEAK 
RATE screen. In this example, the units of measure is g/yr. If your leak source is marked in Oz/yr, then 
multiply that value by a factor of 28.35. For example: If the leak source is marked 0.9 Oz/yr, then enter a 
leak rate of (0.9 x 28.35=) 25.5 g/yr.  
5.2.4 Leak Gas 
 
 
 
 
The Leak Gas setting is set to match the gas contained in the internal leak source as indicated on the tag 
attached to the leak source. Use the Up and Down buttons to highlight the correct gas, and then press 
ENT to select that gas. 
5.2.5 Leak Temp 
 
 
 
 
The Leak Temperature value is set to match the temperature of the internal leak source at its time of 
manufacture as indicated on the tag attached to the leak source. Use the Left and Right buttons to move 
the blinking cursor across the screen to highlight the digit to be modified, and then use the Up and Down 
buttons to modify that digit. Press ENT to save the new Leak Temp value. 
LEAK RATE 
2
5.5 g/yr 
DIGIPOT=197 
IR VOLT=4.2000 
DIGIPOT=197 
IR VOLT=4.2000 
LEAK GAS 
<
 R134A
 
  
 
R401A 
  
R402A 
 < 
R402B 
 
LEAK TEMP 
0
    2
3
.9degC