2.4.2 Vessel 3 to sample cell 
 
The transfer of liquid from Vessel 3 to the sample cell is brought about by the 
difference in pressure between that in the Vessel and a vacuum in the cell.  However, 
in order to control the transfer of liquid the pressure difference should be as small as 
possible but not so low that the liquid starts to boil. 
 
•  Switch the pump to its roughing mode. 
•  Open values V2 and V3 along with tap C and carefully open valve V1 until a 
suitable pressure is obtained (~0.5 bar). 
•  Close tap C and continue to evacuate the system and sample cell by opening 
valves V3, V4, V5 and V8.  Switch the pump to turbo mode. 
•  When the required level of vacuum has been obtained, close valves V4 and V5 
in order to isolate the sample cell. 
•  Opening tap D on Vessel 3 will cause the liquid to move rapidly to fill the 
volume between tap D and V5. 
•  Valve V5 can be opened carefully in order to allow the liquid into the sample 
cell.  The movement of liquid can be monitored using the sight glasses 
positioned either side of the sample cell. 
•  When it is judged that sufficient liquid has been transferred to fill the sample 
cell, close valve V5. 
 
At this point, it must be remembered that the lower manifold contains liquid, which 
must be removed before the system is evacuated. 
 
2.4.3 Recovery of liquid in manifold to Vessel 2 
 
After the transfer of liquid to or from the sample cell, some liquid will remain in the 
pipes of the lower manifold.  This can be removed using the following procedure. 
 
•  Vessel 2 should be under vacuum prior to the loading of the liquid into the 
sample cell.  If it is not, switch the pump to roughing mode and open tap E 
followed by valves V1 and 2.  If Vessel 2 already contains some liquid ensure 
that it is frozen using liquid nitrogen before evacuation is attempted. 
•  When the vessel is evacuated close tap E and cool the glass bulb by raising a 
dewar of liquid nitrogen around it. 
•  Ensure that taps A, B, C, D and E are closed before opening valves V6 and 
V8. 
•  Opening valve V7 will allow any liquid present in the manifold to enter Vessel 
2 where it will freeze. 
•  Close valve V7 and repressurise the system and Vessel 2 by opening valve V9 
and tap E.  Remove the dewar from around the vessel and allow the liquid to 
thaw. 
 
 
 
 - 