3M DL Manuale Utente

Pagina di 168
23    Basic setups 
 
  Navigating 
 
 
Models SE and DL 
 
 
Basic setups 
This chapter covers the basic features and setups of SoundPro SE/DL that 
establish the instrument’s operating conditions. For information about the 
more advanced SoundPro SE/DL features and their setups, see Chapter 4, 
“Advanced setups.” For information on setting-up and measuring add-on 
features (Reverberation, STI-PA, or Curves), please see Chapter 6. 
 
•  
Microphone Settings  
•  
Setting Date-Time 
•  
Setting Measurement type (SLM, 1/1, 1/3, STI-PA) 
•  
Setting up Meter 1/Meter 2 parameters 
•  
Setting up Exceedance levels 
•  
Setting up Logging 
•  
Setting physical settings: backlight “on” time, contrast, choosing a 
language 
•  
Microphone settings 
With Class/Type 1 models, the microphone’s sensitivity, range cap, and 
polarization can be checked and/or changed in the “Signal Input” menu 
(when stopped).  Alternatively, Class/Type 2 models settings are viewable at 
any time the session is stopped but cannot be modified. 
 
The following table displays the required microphone settings. 
 
 
 
 
 
 
Table 3-1:  Required microphone settings 
 
 
Signal Input 
BK4936  QE7052  QE4110 
QE4130 
QE4150  QE4170 
Sensitivity 
a
 
-28.0 
-29.0 
-49.1 
-36.0 
-26.0 
-28.0 
Range Cap (dB) 
140 
140 
160 
140 
140 
140 
Polarization (V) 
200 
200 
200 
200 
 
  CHAPTER