108      Speech Intelligibility  
                 STI-PA results screens
 
 
 
Models SE and DL 
 
 
 
 
 
 
 
 
Figure 6-26:  Captured curves/background noise 
5.  Once all background noise has been captured, refer to, “Applying captured curves” 
below. 
Applying captured curves for STI-PA testing  
To apply the curves during a STI-PA measurement remember to complete one 
and two below before applying the curve.   
   Applying captured curves 
1.  From the Start screen, select View Session and  press     ..  Press Altf key.    
2.  Select one of the captured curves by pressing the 1
st
 softkey (i.e, Cap-1).  
Press On/Off/Esc key. 
3.  From the Start screen, press the 1
st
 softkey continuously until STI-PA is 
selected.  Then, select View Session and the STI-PA screen will open. 
 
 
 
 
 
 
Figure 6-27:  Post CAP-1 applied during STI-PA test 
CAP-1 
Press first 
softkey to assign 
a captured curve 
name (i.e., CAP-
1, CAP-2, CAP-
3, or CAP-4)   
Capture 
Press Capture 
softkey to save 
the background 
noise/SPL. 
Curve indicator 
POST CAP-1 
The selected Captured Curve, 
i.e., CAP-1, will appear on the 
STI-PA screen.  To change the 
Cap-x, repeat steps 1-2 above.