Omega CNI16D Manuale Utente

Pagina di 72
24
Conversion number is a coefficient of conversion between input values
and real full display range (10000 counts shown as 9999). See Table 3.2
below for proper conversion number.
Table 3.2 Conversion Table
RANGE
CONVERSION NUMBER
100 mV
10000 / (100 x 1) = 100
1 V
10000 / (1000 x 1) = 10
10 V
10000 / (1000 x 10) = 1
0 -20 mA
10000 / (20 x 1) = 500
Example =
0 - 1 V = 0 - 100.0
In 1 = 0
Rd 1 = 0
Inp 2 = 9999
Rd 2 = 100.0