Agilent Technologies 24A Manuale Utente

Pagina di 266
4-39
MegaZoom Concepts and Oscilloscope Operation
Using Digital Channels to Probe Circuits
High-Frequency Probe Equivalent Circuit 
The impedance plots for the two models are shown in these figures. By 
comparing the two plots, you can see that both the series tip resistor and the 
cable’s characteristic impedance extend the input impedance significantly. The 
stray tip capacitance, which is generally small (1 pF), sets the final break point 
on the impedance chart.
Impedance versus Frequency for Both Probe Circuit Models 
The logic probes are represented by the high-frequency circuit model shown 
above. They are designed to provide as much series tip resistance as possible. 
Stray tip capacitance to ground is minimized by the proper mechanical design 
of the probe tip assembly. This provides the maximum input impedance at high 
frequencies.
1 pF
7.5 pF
150  
W
250  
W
100 k 
W
100 k 
W
10 k 
W
1 k 
W
100  
W
10  
W
1  
W
10 kHz
100 kHz
1 MHz
10 MHz
100 MHz
1 GHz
High 
Frequency 
Model
Typical 
Model
Frequency
Im
ped
an
ce