 Turn-On  Checkout  53
 Table 3-2. Checking the Current Functions (Output Terminals Shorted) (continued)
 Action
 
 Display
 
 Explanation
 
 Press 
 
 AMPS  0.000
 
 Dis annunciator turns on.
 Press 
 
 
 You have disabled the overcurrent protection circuit. The OCP
annunciator turns off.
 
 Press 
 (
 
)**
 
 
 You have cleared the overcurrent protection circuit. The Prot
annunciator turns off.
 
 Press 
 
 AMPS  1.000
 
 Dis turns off and CC turns on. The output current is restored.
 
 If you have a shorting wire of sufficient capacity,  you may continue testing up to the maximum rated current of the power
supply (see "Performance Specifications"). When finished, go to the next step.
 
 
Press 
 AMPS  0.000
 Dis turns on and output current drops to zero.
 
 Turn off the power supply and remove the short from the output terminals.
 
 **
 is the unlabeled blue key.
 
 
 Checking The Save And Recall Functions (All Models)
 
 The Series 668xA supplies have four nonvolatile memory storage locations (0 through 3). The supplies of all other series
have five (locations 0 through 4). Proceed as follows:
 
■ 
Make certain that the output is on (Dis annunciator is off).
■ 
Set the voltage output to 5 by pressing 
.
■ 
Save this value to location 1 by pressing 
.
■ 
Return the output voltage to 0 by pressing 
 (This step is based on the fact that a newly shipped power
supply has the *RST parameters stored in location 0 (see "Chapter 5 - Front Panel" for more information).
■ 
Press 
 and notice that the output voltage returns to the value stored in location 1.
 
 Determining The GPIB Address (All Models)
 
 When the power supply is turned on, the display shows ADDR n, where n is the power supply GPIB address. Any time you
want to see the address, press 
.
 
 The display will indicate ADDR 5, which is the factory default. If the address has been changed, then a different number
will appear (see “Setting the GPIB Address” in “Chapter 5 - Front Panel”).
 
 
 In Case Of Trouble
 
 Line Fuse
 
 If the power supply appears "dead" with a blank display and the fan not running, first check your power source to be certain
line voltage is being supplied to the power supply. If the power source is normal, the power supply line fuse may be
defective. (On Series 668xA supplies, if the Check Fuses LED is blinking, then one or two of the line fuses are open.) If
the supply has a defective fuse, replace it only once. If it fails again, investigate the reason for the failure. Proceed as
follows: