 Calibration
 
 
 Introduction
 
 The power supply may be calibrated either from the front panel or from a controller over the GPIB. The procedures given
here apply to all models.
 
 
Important 
These instructions do not include verification procedures. If you need to perform verification as a
erequisite to or as part of your calibration procedure, see “Appendix B - Verification”.
 
 
 Equipment Required
 
 The equipment listed in Table A-1, or equivalent, is required for calibration.
 
 Table A-1. Equipment Required For Calibration
 Equipment
 Characteristics
 Recommended Model
 Voltmeter
 D-c accuracy 0.005%, 6 digits
 Agilent 3456A or 3458A
 Shunt resistor
 
 
      Agilent 6641A, 51A, 52A
 100 A, 0.01 
Ω
, 0.04%, 100 W
 Guildline 9230/100
      Agilent 6642A, 43A, 44A, 45A,
                   6643A, 54A, 55A
 15 A, 0.1 
Ω
, 0.04%, 25 W
 
 Guildline  9230/15
 
      Agilent  6671A
 300 A, 0.001 
Ω
, 0.04%,100 W
 Guildline 9230/300
      Agilent 6672A, 73A, 74A, 75A
 300 A, 0.001 
Ω
, 0.04%, 100 W
 Guildline 9230/300
      Agilent 6680A, 81A
 1000 A, 0.1 m
Ω
, 0.05%
 Burster 1280S
      Agilent 6682A, 83A, 84A
 300 A, 0.1 m
Ω
, 0.05%
 
 Guildline  9230/300
 
 For Calibration Over the GPIB
 GPIB  Controller
 
 HP Vectra (or IBM compatible) with GPIB Interface, or HP BASIC series
 
 
 General Procedure
 
 Because the power supply output must be enabled during calibration, voltages or currents
hazardous to personnel and/or damaging to equipment can appear at the output terminals.
 
 
 
 ENERGY HAZARD.  Series 668xA supplies can provide more than 240 VA at more than 2 V. If the
output connections touch, severe arcing may occur resulting in burns, ignition or welding of parts. Do
not attempt to make connections to live output circuits.
 
 Parameters Calibrated
 
 The following parameters may be calibrated:
• 
Output voltage.
• 
Output voltage readback.
• 
Overvoltage protection (OVP).