 Calibration  97
 Table A-2. Typical Front Panel Calibration Procedure (continued)
 Action
 
 Display Response
 
 Calibrating Current Monitor (I
M
) (Series 668xA Only)
 
 If you perform this calibration, then you must recalibrate the current output.
 1. Make certain the appropriate shunt resistor (see Table A-1) is the only load on the power
supply.
 2. Select IMN calibration by pressing 
 
    If the power supply is not in CC mode, an error occurs.
4
 3. Wait for DVM reading to stabilize. Then read DVM and compute the current value
     (DVM reading 
÷
 shunt resistance).
 4. Use Entry keypad to enter the current value.
  Note: If the entered value is not within acceptable range, an error occurs.
     Wait for the power supply to compute the new current calibration constants, which will be
     stored in RAM.
 If the constant is not within acceptable range, an error occurs.
 
 
 (Meter  mode)
 
 IMON  CAL
 WRONG MODE
 (Meter  mode)
 
 (Meter  mode)
 CAL  ERROR
 CAL  COMPLETE
 
 CAL ERROR
 
 
4.
  Program the output voltage to 10% of its rated output*
 *See applicable Output Ratings in “Chapter 1- General Information”
 
 
 
 Recovering From Calibration Problems
 
 You can encounter serious calibration problems if you cannot determine a calibration password that has been changed or the
power supply is severely out of calibration. There are jumpers inside the power supply that permit the calibration password
to be defeated and allow the original factory calibration constants to be restored. These jumpers are explained in the Service
Manual.
 
 Calibration Error Messages
 
 Error messages that can occur during calibration are shown in Table A-3.
 
 Table A-3. GPIB Calibration Error Messages
 Error
 No.
 Meaning
 
 
 Error
 No.
 Meaning
 
 1
 CAL jumper prevents calibration
1
 
 6
 Wrong CAL command sequence
 2
 CAL password is incorrect
 
 7
 Incorrect state (CV/CC) for this command
 3
 CAL mode is not enabled
 
 
 
 4
 Incorrect computed readback constants
 
 
 
 5
 Incorrect  computed  programming
constants
 
 
1
 This is a hardware disable. See the power supply
Service Manual.