Computer Interface, Commands, and Data Collection 
07506C DCN6727 
71 
 
 
Example of data record: 
 
TSI CPC DATA VERSION 3 
1268228469,2010/3/10,13:41:09 
60 
1.00,120 
 Model 651 Ver 1.00 S/N 123456 
"Date","Time","Concentration","Count"," Live-
Time","Blank","Abs Press"," Analog In","Pulse Height"," Pulse 
STD","Status Flags" 
2010/3/10,13:41:57,2.15e4,2522183,58.62,,970,0.00,567,600,0 
2010/3/10,13:41:57,2.32e4,2719488,58.51,,970,0.00,607,595,0 
2010/3/10,13:42:57,2.15e4,2530791,58.62,,970,0.00,587,609,0 
2010/3/10,13:43:57,2.13e4,2505886,58.63,,970,0.00,581,615,0 
 
Data fields include: 
  Date 
  Time 
  Particle Concentration (#/cm
3
  Raw Particle Counts 
  Live-time (seconds) 
  Comma (reserved field) 
  Absoluter Pressure (mbars) 
  Analog Input (V) 
  Pulse Height average (mV) 
  Pulse Height Standard Deviation 
  Status flags. 
 
You can select the Sample Time (the period over which data is 
collected and reported) from one of the following choices: 1, 2, 3, 4, 
5, 6, 10, 12, 15, 20, 30, or 60 seconds. Data is collected internally 
10 times/second and is averaged over the selected sample time. 
The average is displayed on the graph and can be saved to the flash 
drive. Data is saved to the flash drive every 10 seconds, or at the 
rate of the data averaging period if it is longer than 10 seconds. The 
data averaging period is the same as the Sample Time setting. Once 
the data is being logged to the Flash drive, the Sample Time setting 
cannot be changed. 
 
IMPORTANT
 
If power is lost at any time, the instrument should continue data logging 
when the power is returned. The data files created will have the extension 
.rdt
 so that the previous files are not overwritten.