Main Menu
(continued)
_ CALIBRATION
Select CALIBRATION in the “SYSTEM CONFIG” menu to open the following menus.
*1 “EXECUTING” is displayed during “AUTO CALIBRATION” and “COMPLETE” appears when calibration ends.
“INCOMPLETE” appears if calibration could not be completed.
*2 When “RESET” ends, “COMPLETE” appears.
1. Press [
,
] in the “SYSTEM CONFIG” menu,
select the [CALIBRATION] submenu and press
[ENTER].
[ENTER].
2. Press [
,
], select the [AUTO CALIBRATION]
submenu and press [ENTER].
3. Select “YES” and press [ENTER].
This starts calibration.
xxxxxxxx indicates that one of the following
messages with the meaning listed below appears.
messages with the meaning listed below appears.
EXECUTING: Operation in progress
COMPLETE: Operation completed
INCOMPLETE: Operation did not complete.
(Check RS-232C cable connection.)
COMPLETE: Operation completed
INCOMPLETE: Operation did not complete.
(Check RS-232C cable connection.)
1. Press [
,
] in the “SYSTEM CONFIG” menu,
select the [CALIBRATION] submenu and press
[ENTER].
[ENTER].
2. Press [
,
], select the [RESET] submenu and
press [ENTER].
3. Select “YES” and press [ENTER].
This resets the setting of this monitor.
xxxxxxxx indicates that the following message with
the meaning listed below appears.
the meaning listed below appears.
COMPLETE: Operation completed
Sub menu
Setting
Description
AUTO CALIBRATION
*1
Connect a CA-210 color analyzer and use this submenu to make a
calibration.
Select “AUTO CALIBRATION” and select “YES” in the
confirmation screen that appears to start calibration.
calibration.
Select “AUTO CALIBRATION” and select “YES” in the
confirmation screen that appears to start calibration.
RESET
*2
Returns calibration data to their factory defaults.
Select “RESET” and select “YES” in the confirmation screen that
appears to return calibrated values to their factory defaults.
Select “RESET” and select “YES” in the confirmation screen that
appears to return calibrated values to their factory defaults.
Performing AUTO CALIBRATION
[AUTO CALIBRATION]
[AUTO CALIBRATION]
RESET operation
[RESET]
[RESET]