Parameter Settings Transfer
Setting up your Monitor   3-61
S
e
tt
ing up y
our
 
M
onit
o
r  
Parameter Settings Transfer
Parameter modules which are labeled on the front with a small “T” in the 
upper right hand corner can be transported from one rack to another (or 
from one Philips system to another) and still keep their parameter 
settings.
Note—
There are a few modules that may or may not be marked with a “T” 
that DO transfer parameter settings. They are: M1006A, M1016A, M1006A-
#CO1,    M1029A, M1006B, M1032A, M1012A, M1012A #C10, M1034A. Any 
other module without a “T” does not transfer parameter settings. 
The settings (such as alarm limits) are stored inside the module. You 
could even unplug a satellite rack full of these modules from one monitor, 
and plug it into another monitor without losing the parameter settings. 
This behavior permits fast and easy transport, and is called “Parameter 
Settings Transfer”.
Note—
The settings that are stored in each module are detailed in the 
relevant module chapters of this guide. 
For Parameter Settings Transfer to work on your monitor, it must be set 
ON in a special Service Mode, either by your biomedical engineering 
department or the Philips service engineer.
After making changes to parameter settings, wait about 10 seconds until 
the new settings are stored in the relevant module before removing the 
module.
After plugging in a module, wait about 10 seconds to let the monitor read 
the settings before opening the Task Window.
After switching on the system or going into Monitor mode from Config 
mode, the first 30 seconds are used to store settings from the system to 
the modules. Modules plugged in after this initial 30 second period are 
capable of keeping their settings.
You can change parameter settings either:
in the parameter Task Window,