Philips V26CT Manuale Utente
Controls and Connectors
11-34 Monitor Installation and Patient Safety
M
onitor
Ins
ta
lla
ti
on
a
nd Pa
ti
e
n
t Sa
fe
ty
The V24 and
V26
Connectors
V26
Connectors
The monitor has several connectors which are shown below.
Do not apply a voltage of more than +12V to any of the pins on the
defibrillator sync, parameter module rack, Philips monitoring network
(SDN) or equipotential ground connectors.
defibrillator sync, parameter module rack, Philips monitoring network
(SDN) or equipotential ground connectors.
The secondary ground complies with IEC 601-1-1 as shown on page 10-4.
The connectors on the monitor are:
1. AC Power Connector; this connector is used to connect the
monitor to an AC power source.
Note—
If you have a Philips M1026A Anesthetic Gas Module (AGM), do
not daisy-chain the AC power through the AGM. This configuration may
degrade RFI performance.
degrade RFI performance.
2. Secondary Ground Screw; this is a posidrive screw to connect the
monitor chassis to earth ground per IEC 601-1-1 standard.
3. Equipotential Ground Post, this connects the monitor to the
equipotential ground system.
AC LINE INPUT
RS232 CONNECTOR
(Rear Housing with
Serial Prefix <3608)
PROTECTIVE
RS232 CONNECTOR
(Rear Housing with
Serial Prefix >=3608)
EQUIPOTENTIAL GROUNDING
DEFIB
CONNECTOR
RACK
CONNECTOR
marge3.tif
EARTH GROUND
PHILIPS MONITORING NETWORK (SDN)
SINGLE CONNECTOR