f)  Using a 0-5VDC Input / Output Power Adapter Package. 
 
An optional 0-5VDC Input / Output Power Adapter Package is available for 
use with the FMA 3200/3400 Series. This consists of a power source 
(115VAC or 230VAC), a connection hub and two cable assemblies with pig-
tail (soldered wire) ends. This should be assembled as shown in the 
following diagram.  
 
Assembling a FMA 3215PW Power Adapter Package 
(the FMA 3223PW Power Adapter Package is similar) 
 
 
 
 
The RED connector should be inserted in the RED socket on the 
connection hub. The WHITE connector should be inserted in the WHITE 
socket on the connection hub. 
 
The cable with a RED connector provides the input signal. The RED wire 
of this cable should be connected to the positive terminal of the voltage 
source. The (0-5VDC) voltage control signal should be supplied from a low 
impedance source.  The bare wire of this cable assembly is the input 
signal negative (ground) and should be connected to the negative 
(Ground) terminal of the voltage source. 
 
The cable with a WHITE connector provides the signal output. The 
WHITE wire should be connected to the positive terminal of the display, 
data acquisition system or voltmeter with an impedance of greater than 
2.5 kΩ (kilo ohms). The bare wire of this cable assembly is the signal 
negative (ground) and should be connected to the negative (Ground) 
terminal of the display, data acquisition system or voltmeter. 
 
 
Caution: Avoid high voltage static discharges to the input signal 
connection. Do not short the output signal wires or allow them to 
contact the power wires at any time. DAMAGE WILL RESULT!