Out    105.8 mbar  
9.3.13
Installation parameter procedure
  To enter this procedure press the <ENTER> + <ON> keys
  The display will indicate as shown below (the  indicated numerical value  is only  an  example).
CAL?  65388   N=0  
  To exit from the procedure, in any of the following steps switch the instrument -Off-.
9.3.13.1
Firmware version code - Serial number
From the above step of the procedure it is possible to view the software version code .
 Press 
the 
<IN/OUT>
 key to obtain the following indication:
Ver2.000  #  65535 
.
The reading on the display indicates that the instrument is equipped with a memory release code 2.000.
The above information is extremely useful to understand the update  status   of the instrument and to simplify information
exchange with OMEGA engineers during repair or service operations.
The second number on the right side of the display is the Serial Number of the instrument.
  Press any key  to exit  the procedure with the following indication:
CAL?  65388   N=0   
   Switch the instrument -OFF- to end  the  procedure.
9.3.13.2
External Rj compensation
To enter the External Rj compensation set-up start with the instrument switched -Off- and press <ENTER> + <ON> keys
to obtain the following indication:
CAL?  65388   N=0  
  Press  the  <AUTORAMP> key to obtain:
RJ :     0.0°C ext 
  Set, with <▲> and <▼> keys, the  temperature of the external Rj compensation (adjustable from -50°C to +100°C).
 Press 
the 
<AUTORAMP>
 key to memory load  the new temperature value of reference junction compensation.
  To end this   procedure switch the instrument -Off-